ACCUMAX SMART DISPENSER
Special Teflon seal for long-lasting, precise and smooth operation
Comfortable finger rest for both right and left hand use
On-site cleaning, repair and calibration without the use of special tools
New tip cone design for compatibility with a wide range of tips
Locking mechanism to prevent accidental volume changes during pipetting
Soft ejector for smooth one-handed extraction and pollution prevention.
Lightweight design for continued user-friendliness during long-term use
Optional: A tripod carousel with 6 single-channel dispensers.
SVA-100
Volume range: 0.5–10 µl
Increments: 0.02 µl
Test Volume / Inaccuracy (±%) / Inaccuracy [µl] / Error (±%) / Error [µl]:
– 1 ml: 2.5% / 0.03 µl / 1.8% / 0.02 µl
– 5 µl: 2% / 0.10 µl / 1% / 0.05 µl
– 10 µl: 1% / 0.10 µl / 0.6% / 0.06 µl
SVA-125
Volume range: 0.1–2.5 µl
Increments: 0.002 µl
– 0.25 µl: 12% / 0.03 µl / 6% / 0.02 µl
– 1.25 µl: 3% / 0.04 µl / 3% / 0.04 µl
– 2.5 µl: 2.5% / 0.06 µl / 1.6% / 0.04 µl
SVA-200
Volume range: 5-50 µl
Increments: 0.1 µl
– 5 µl: 2% / 0.10 µl / 2% / 0.10 µl
– 25 µl: 0.8% / 0.20 µl / 0.4% / 0.10 µl
– 50 µl: 0.6% / 0.30 µl / 0.3% / 0.15 µl
SVA-500
Volume range: 10-100 µl
Increments: 0.1 µl
– 10 µl: 3% / 0.30 µl / 1% / 0.10 µl
– 50 µl: 1% / 0.50 µl / 0.5% / 0.25 µl
– 100 µl: 0.8% / 0.80 µl / 0.2% / 0.20 µl
SVA-600
Volume range: 100-1000 µl
Increments: 0.1 µl
– 100 µl: 3% / 3.00 µl / 0.6% / 0.60 µl
– 500 µl: 1% / 5.00 µl / 0.4% / 2.00 µl
– 1000 µl: 0.6% / 6.00 µl / 0.2% / 2.00 µl
SVA-700
Volume range: 2-20 µl
Increments: 0.02 µl
– 2 µl: 3% / 0.06 µl / 2% / 0.04 µl
– 10 µl: 1.5% / 0.15 µl / 1% / 0.10 µl
– 20 µl: 0.9% / 0.18 µl / 0.4% / 0.08 µl
SVA-800
Volume range: 20-200 µl
Increments: 0.2 µl
– 20 µl: 2.5% / 0.50 µl / 0.7% / 0.14 µl
– 100 µl: 0.7% / 0.70 µl / 0.3% / 0.30 µl
– 200 µl: 0.6% / 1.20 µl / 0.2% / 0.40 µl
SVA-905
Volume range: 500-5000 µl
Increments: 10 µl
– 500 µl: 2.4% / 12.00 µl / 0.6% / 3.00 µl
– 2500 µl: 1.2% / 30.00 µl / 0.25% / 6.25 µl
– 5000 µl: 0.6% / 30.00 µl / 0.2% / 10.00 µl
SVA-1000
Volume range: 1000-10000 µl
Increments: 20 µl
– 1000 µl: 3% / 30.00 µl / 0.6% / 6.00 µl
– 5000 µl: 0.8% / 40.00 µl / 0.3% / 15.00 µl
– 10000 µl: 0.6% / 60.00 µl / 0.2% / 20.00 µl
SFA-2.5
Volume Range: 2.5 µl
Test volume: 2.5 µl
Inaccuracy (±): 2% / 0.05 µl
Error (±): 1.6% / 0.04 µl
SFA-0005
Volume range: 5 µl
Test volume: 5 µl
Inaccuracy: 1.3% / 0.07 µl
Error rate: 1.2% / 0.06 µl
SFA-0010
Volume range: 10 µl
Test volume: 10 µl
Inaccuracy: 1.2% / 0.12 µl
Error rate: 0.8% / 0.08 µl
SFA-0020
Volume range: 20 µl
Test volume: 20 µl
Inaccuracy: 1% / 0.20 µl
Error rate: 0.5% / 0.10 µl
SFA-0025
Volume range: 25 µl
Test volume: 25 µl
Inaccuracy: 1% / 0.25 µl
Error rate: 0.3% / 0.08 µl
SFA-0050
Volume range: 50 µl
Test volume: 50 µl
Inaccuracy: 0.7% / 0.35 µl
Error rate: 0.3% / 0.15 µl
SFA-0100
Volume range: 100 µl
Test volume: 100 ml
Inaccuracy: 0.6% / 0.60 µl
Error rate: 0.2% / 0.20 µl
SFA-0200
Volume range: 200 µl
Test volume: 200 ml
Inaccuracy: 0.6% / 1.20 µl
Error rate: 0.2% / 0.40 µl
SFA-0250
Volume range: 250 µl
Test volume: 250 ml
Inaccuracy: 0.6% / 1.50 µl
Error rate: 0.2% / 0.50 µl
SFA-0500
Volume range: 500 µl
Test volume: 500 ml
Inaccuracy: 0.6% / 3.00 µl
Error rate: 0.2% / 1.00 µl
SFA-1000
Volume range: 1000 µl
Test volume: 1000 µl
Inaccuracy: 0.6% / 6.00 µl
Error rate: 0.2% / 2.00 µl
SFA-2000
Volume range: 2000 µl
Test volume: 2000 µl
Inaccuracy: 0.3% / 6.00 µl
Error: 0.15% / 3.00 µl
SFA-5000
Volume range: 5000 µl
Test volume: 5000 µl
Inaccuracy: 0.3% / 15.00 µl
Accuracy: 0.15% / 7.50 µl
SFA-10000
Volume range: 10000 µl
Test volume: 10000 µl
Inaccuracy: 0.6% / 60.00 µl
Error rate: 0.2% / 20.00 µl
SMA8-010
Volume range: 0.5–10 µl
Increments: 0.02 µl
Test Volume / Inaccuracy (±%) / Inaccuracy [µl] / Error (±%) / Error [µl]:
– 10 µl: 2% / 0.20 µl / 1% / 0.10 µl
– 5 µl: 4% / 0.20 µl / 2% / 0.10 µl
– 1 µl: 8% / 0.08 µl / 5% / 0.05 µl
SMA8-020
Volume range: 2-20 µl
Increments: 0.02 µl
– 20 µl: 2% / 0.40 µl / 1.6% / 0.32 µl
– 10 µl: 3% / 0.30 µl / 2% / 0.20 µl
– 2 µl: 7% / 0.14 µl / 3% / 0.06 µl
SMA8-050
Volume range: 5-50 µl
Increments: 0.1 µl
– 50 µl: 1% / 0.50 µl / 0.7% / 0.35 µl
– 25 µl: 1.5% / 0.38 µl / 1% / 0.25 µl
– 5 µl: 3% / 0.15 µl / 2% / 0.10 µl
SMA8-100
Volume range: 10-100 µl
Increments: 0.1 µl
– 100 µl: 0.8% / 0.80 µl / 0.3% / 0.30 µl
– 50 µl: 1% / 0.50 µl / 0.8% / 0.40 µl
– 10 µl: 3% / 0.30 µl / 2% / 0.20 µl
SMA8-200
Volume range: 20-200 µl
Increments: 0.2 µl
– 200 µl: 0.7% / 1.40 µl / 0.25% / 0.50 µl
– 100 µl: 1% / 1.00 µl / 0.4% / 0.40 µl
– 20 µl: 5% / 1.00 µl / 1.4% / 0.28 µl
SMA8-300
Volume range: 30-300 µl
Increments: 0.2 µl
– 300 µl: 0.6% / 1.80 µl / 0.3% / 0.90 µl
– 150 µl: 1% / 1.50 µl / 0.5% / 0.75 µl
– 30 µl: 3% / 0.90 µl / 1% / 0.30 µl
SMA12-010
Volume range: 0.5–10 µl
Increments: 0.02 µl
Test Volume / Inaccuracy (±%) / Inaccuracy [µl] / Error (±%) / Error [µl]:
– 10 µl: 2% / 0.20 µl / 1% / 0.10 µl
– 5 µl: 4% / 0.20 µl / 2% / 0.10 µl
– 1 µl: 8% / 0.08 µl / 5% / 0.05 µl
SMA12-020
Volume range: 2-20 µl
Increments: 0.02 µl
– 20 µl: 2% / 0.40 µl / 1.6% / 0.32 µl
– 10 µl: 3% / 0.30 µl / 2% / 0.20 µl
– 2 µl: 7% / 0.14 µl / 3% / 0.06 µl
SMA12-050
Volume range: 5-50 µl
Increments: 0.1 µl
– 50 µl: 1% / 0.50 µl / 0.7% / 0.35 µl
– 25 µl: 1.5% / 0.38 µl / 1% / 0.25 µl
– 5 µl: 3% / 0.15 µl / 2% / 0.10 µl
SMA12-100
Volume range: 10-100 µl
Increments: 0.1 µl
– 100 µl: 0.8% / 0.80 µl / 0.3% / 0.30 µl
– 50 µl: 1% / 0.50 µl / 0.8% / 0.40 µl
– 10 µl: 3% / 0.30 µl / 2% / 0.20 µl
SMA12-200
Volume range: 20-200 µl
Increments: 0.2 µl
– 200 µl: 0.7% / 1.40 µl / 0.25% / 0.50 µl
– 100 µl: 1% / 1.00 µl / 0.4% / 0.40 µl
– 20 µl: 5% / 1.00 µl / 1.4% / 0.28 µl
SMA12-300
Volume range: 30-300 µl
Increments: 0.2 µl
– 300 µl: 0.6% / 1.80 µl / 0.3% / 0.90 µl
– 150 µl: 1% / 1.50 µl / 0.5% / 0.75 µl
– 30 µl: 3% / 0.90 µl / 1% / 0.30 µl